**Note** The elevation correction chart supports new style Meth-
ane ODORATORS only. Instruments manufactured
before July 1997, that have not been upgraded or cali-
brated to a methane reference other than 1.00% will not
use the elevation correction charts. Consult the factory
for using older style ODORATORS at elevation. To
identify an ODORATOR’S style or to tell if an upgrade
has been performed, check the top panel to see if there
is a snap-in bezel or a clear window in the overlay cov-
ering the display (“LCD”). Older style ODORATORS
have a snap-in bezel over the LCD. ODORATORS not
calibrated for 1.00% methane have correction charts
for that gas or reference level and can be identied by
the chart as the reading and actual values will be equal
at the reference level.